[Recalled] Stryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, o
The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.
Recall details
| Product | Stryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The |
|---|---|
| Brand / manufacturer | Stryker Medical Division of Stryker Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1862-2012 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 17 May 2012 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Stryker Medical Division of Stryker Corporation
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.
- What Country Makes Vitamins And Supplements
- Top Recalled Products by Category and How to Stay Safe When Shopping Online
- Best Exercise Bike for Home Use: Upright, Recumbent, or Spin?
- Is a Product Fake if Its Barcode Starts With 540 to 549? (Belgium & Luxembourg)
- Healthier Cereal Choices: A Nutritional Comparison Guide
- English Literature for Beginners: A Comprehensive Guide