ProductGuru
AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →

[Recalled] GoBed+ . Model FL20E Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver

recall fda device united-states

Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.

Recall details

ProductGoBed+ . Model FL20E Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices li
Brand / manufacturerStryker Medical Division of Stryker Corporation
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-1859-2012
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date17 May 2012

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Stryker Medical Division of Stryker Corporation

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.

Shop Recalled GoBed Model FL20E
Search on Amazon ›Find on eBay ›
As an Amazon Associate and eBay Partner, we may earn from qualifying purchases.
Related articles

Explore by topic

Looking up a product? Use ProductGuru — free barcode & product database with 35M+ products. Search by EAN / UPC barcode.
Was this page helpful?