[Recalled] Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM1.S Rx Only, Manufacturer GMS- Gesellschaft fur medizinische, Sondentechnik mbH Dofstrasse 2 24247 Melkendorf, Germany Distributed by Integ
Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.
Recall details
| Product | Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM1.S Rx Only, Manufacturer GMS- Gesellschaft fur medizinische, Sondentechnik mbH Dofstrasse 2 24247 Melkendorf, Germany Distributed by Integra Neurosciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Used |
|---|---|
| Brand / manufacturer | Integra LifeSciences Corp. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0059-2014 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 11 June 2013 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Integra LifeSciences Corp.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.