[Recalled] Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspirat |
|---|---|
| Brand / manufacturer | Integra LifeSciences Corp. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0694-2018 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 26 September 2017 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.