Recall details
| Product | AXIOM Artis dBC, Model Number 7728392 |
|---|---|
| Brand / manufacturer | Siemens Medical Solutions USA, Inc |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2027-2019 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 12 June 2019 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.