Recall details

ProductSiemens Cios Alpha VA30-mobile X-Ray system, Material # 11105200 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Brand / manufacturerSiemens Medical Solutions USA, Inc
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-1958-2020
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date20 April 2020
EAN / barcode11105200

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Siemens Medical Solutions USA, Inc

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.

Shop Recalled Siemens Cios Alpha
Search on Amazon ›Find on eBay ›
As an Amazon Associate and eBay Partner, we may earn from qualifying purchases.
Related articles

Explore by topic

Looking up a product? Use ProductGuru — free barcode & product database with 35M+ products. Search by EAN / UPC barcode.
Was this page helpful?