[Recalled] Artis Q zeego, Model Number 10848283 — If the coolant level in the cooling circuit drops below a certain level, this m… | FDA 2022-02-23
The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.
Recall details
| Product | Artis Q zeego, Model Number 10848283 |
|---|---|
| Brand / manufacturer | Siemens Medical Solutions USA, Inc |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0977-2022 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 23 February 2022 |
| EAN / barcode | 4056869010007 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Siemens Medical Solutions USA, Inc
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.