The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas

Safety check

🟡 Open recall on file · 2024-10-22 · Australia
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, u...
Source: US FDA - Medical Devices · last fetched 2026-05-31 · Verify on US FDA → · Full notice details
🟡 Open recall on file · 2023-11-28 · Australia
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
Source: US FDA - Medical Devices · last fetched 2026-05-31 · Verify on US FDA → · Full notice details
🟢 Historical recall · 2023-05-31 · Australia
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Source: US FDA - Medical Devices · last fetched 2026-05-31 · Verify on US FDA → · Full notice details

ProductGuru indexes public regulatory data; we are not the issuing authority. For the official record, follow each "Verify on ..." link. If this product is medication, ask a pharmacist or doctor.

Product Details

Barcode (EAN/GTIN)9357123000037
Product nameThe device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas
Brand / ManufacturerMICRO-X LIMITED
CategoryMedical Device
GS1 Prefix93 — registered in Australia
Barcode typeEAN-13 / GTIN-13

Recorded specifications

No measured specifications are on file for barcode 9357123000037 yet. Barcodes that people look up get enriched over time.

About EAN Barcode 9357123000037

Barcode 9357123000037 is a EAN-13 (GTIN-13) identifier in the GS1 global system, identifying The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas.

Frequently Asked Questions

What product has barcode 9357123000037?

The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas by MICRO-X LIMITED

Is there a recall on barcode 9357123000037?

See the safety check above — we compare every barcode against EU Safety Gate, US FDA, US CPSC and Health Canada.

What country is barcode 9357123000037 from?

The prefix 93 was issued by GS1 Australia (Australia). That is where the brand owner registered the barcode — not necessarily where the product is made.