🟡 Open recall on file · 2025-05-15 · United States
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
Source: US FDA - Medical Devices · last fetched 2026-05-31 · Verify on US FDA → · Full notice details
🟢 Historical recall · 2023-06-19 · United States
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cance...
Source: US FDA - Medical Devices · last fetched 2026-05-31 · Verify on US FDA → · Full notice details
🟢 Historical recall · 2023-06-19 · United States
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cance...
Source: US FDA - Medical Devices · last fetched 2026-05-31 · Verify on US FDA → · Full notice details