[Recalled] XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen an
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distribu |
|---|---|
| Brand / manufacturer | Elekta, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0780-2014 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 25 November 2013 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Elekta, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.