[Recalled] WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems. — Boston Scientific has identified
The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.
Recall details
| Product | WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems. |
|---|---|
| Brand / manufacturer | Boston Scientific Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2429-2025 |
| Country of notice | United States |
| Risk level | Class I |
| Hazard type | recall |
| Notice date | 29 July 2025 |
| EAN / barcode | 00191506013820 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
This Class I recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Stop using the product immediately and follow the authority's remedy (return, repair, or disposal).
Read the full official notice →