[Recalled] VITROS Chemistry Products Calibrator Kit 4-(Calibrator values for TIBC only). Used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems Product Code: 1204668 Impacte
The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.
Recall details
| Product | VITROS Chemistry Products Calibrator Kit 4-(Calibrator values for TIBC only). Used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems Product Code: 1204668 Impacted Products: A. VITROS Chemistry Products TIBC Kit Product No: 7541 (U |
|---|---|
| Brand / manufacturer | Ortho-Clinical Diagnostics, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1295-2021 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 18 February 2021 |
| EAN / barcode | 10758750008537 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Potential total iron-binding capacity (TIBC) results. The VITROS TIBC performance shift will start with the newest Generation (GEN) of VITROS Fe Slides (GEN 26), the potential shift in TIBC patient values of GEN 26 and above will be approximately 80 ug/dL lower across the measuring range than other available GENS of VITROS Fe Slides (GENs 22-25).
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Ortho-Clinical Diagnostics, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.