[Recalled] Trilogy EVO Ventilator, 02 - EU - Model Number: EU2100X15B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Puls
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Trilogy EVO Ventilator, 02 - EU - Model Number: EU2100X15B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories. |
|---|---|
| Brand / manufacturer | Philips Respironics, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0953-2020 |
| Country of notice | United States |
| Risk level | Class I |
| Hazard type | recall |
| Notice date | 19 November 2019 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.
This Class I recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Stop using the product immediately and follow the authority's remedy (return, repair, or disposal).
Read the full official notice →
Related recalls from Philips Respironics, Inc.