[Recalled] Trilogy Acetabular Shells with Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage. — Some units of Trilogy shells manufacture
A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.
Recall details
| Product | Trilogy Acetabular Shells with Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage. |
|---|---|
| Brand / manufacturer | Zimmer, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0775-2013 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 14 September 2012 |
| EAN / barcode | 61567730 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →