[Recalled] Torque Limiting Screwdriver: Greatbatch Medical, Torque Limiting Driver, Non Sterile, Rx Only. PRECIMED, Non Sterile. Zimmer, Non Sterile, Rx Only. Screwdriver: Customer Number/Model: 804-03-038 T11623, T14001 T
Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.
Recall details
| Product | Torque Limiting Screwdriver: Greatbatch Medical, Torque Limiting Driver, Non Sterile, Rx Only. PRECIMED, Non Sterile. Zimmer, Non Sterile, Rx Only. Screwdriver: Customer Number/Model: 804-03-038 T11623, T14001 T14001, MTO35173507 T14378, MTO375173506 T14379, MTO35173505 T1438 |
|---|---|
| Brand / manufacturer | Greatbatch Medical |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1883-2016 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 26 April 2016 |
| EAN / barcode | 2743160001 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →