[Recalled] Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used for the percutaneous introduction of various types of cardiovascular catheters and guidew
A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.
Recall details
| Product | Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal p |
|---|---|
| Brand / manufacturer | Boston Scientific Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2106-2025 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 10 June 2025 |
| EAN / barcode | 00685447005391 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Boston Scientific Corporation
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.