[Recalled] The Access hLH assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of luteinizing hormone (LH) levels in human serum and plasma using the Access Immunoassay Systems. The Access hLH
The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.
Recall details
| Product | The Access hLH assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of luteinizing hormone (LH) levels in human serum and plasma using the Access Immunoassay Systems. The Access hLH Calibrators are intended to calibrate the Access hLH assay for the qua |
|---|---|
| Brand / manufacturer | Beckman Coulter Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0572-2013 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 05 November 2012 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Beckman Coulter Inc. (BEC) has demonstrated that real time stability testing of hLH Calibrator of certain lots have failed the 13-month and 14-month time points. Per BEC's real time stability procedure, in order to verify expiration dating, one additional month is needed beyond the claimed expiration period. Therefore, these lots did not meet BEC's procedural requirements for a 12-month shelf-lif
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Beckman Coulter Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.