ProductGuru
AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →

[Recalled] The ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctB

recall fda device united-states

A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.

Recall details

ProductThe ABL software version 6.12 for the ABL800 and ABL700 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions
Brand / manufacturerRadiometer America Inc
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-2163-2012
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date05 July 2012

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Radiometer America Inc

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.

Shop Recalled ABL software version
Search on Amazon ›Find on eBay ›
As an Amazon Associate and eBay Partner, we may earn from qualifying purchases.

Explore by topic

Looking up a product? Use ProductGuru — free barcode & product database with 35M+ products. Search by EAN / UPC barcode.
Was this page helpful?