[Recalled] Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. — Malfunctio
Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.
Recall details
| Product | Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. |
|---|---|
| Brand / manufacturer | Terumo Cardiovascular Systems Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1116-2015 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 06 February 2015 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.