[Recalled] TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: For in vitro coagulation studies — Software defect in TEG Manager impacts the displayed alert for out of range
The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.
Recall details
| Product | TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: For in vitro coagulation studies |
|---|---|
| Brand / manufacturer | Haemonetics Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1617-2020 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 19 February 2020 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →