Recall details
| Product | TDX SP Power Wheelchair, TDX SR Power Wheelchair, TDX SI Power Wheelchair. Model Number(s): TDXSI, TDXSI-2, TDXSI-2-S, TDXSI-CG, TDXSI-HD, TDXSI-HD-S, TDXSIV, TDXSIV-2, TDXSIV-2-S, TDXSIV-HD, TDXSIV-HD-S, TDXSP, TDXSPBASE, TDXSP-CG, TDXSPEURO, TDXSP-MCG, TDXSP-MCG-GT, TDXSR, TDXSR-CG, TDXSR-CG-HD, T |
|---|---|
| Brand / manufacturer | Invacare Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1905-2012 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 06 April 2012 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Invacare Corporation decided to recall the product because of a potential risk of fire and serious injury. The post fuse assemblies could potentially have a short between the positive ring terminal and the terminal battery bracket which would create an alternate current path around the fuse.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Invacare Corporation
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.