[Recalled] System 1 Base 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures,
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | System 1 Base 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in th |
|---|---|
| Brand / manufacturer | Terumo Cardiovascular Systems Corporation |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1969-2012 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 25 June 2012 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump occlusion malfunctions that could result in the inability to use the pump and would then require replacement with back-up equipment. Occlusion malfunctions include stiff/difficult to adjust occlusion, loss of occlusion setting, unequal roller-to-roller occlusion and inability to adjust the occlusion. If it is
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Terumo Cardiovascular Systems Corporation
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.
- What is 1-Hexanesulfonic acid, 1,1,2,2,3,3,4,4,5,5,6,6,6-tridecafluoro-, cesium salt (1:1) (CAS 92011-17-1)? Safety, Regulation & EU/US/Canada Perspective | ProductGuru
- Home Inspection: What Buyers Need to Know
- Was ist 1-bromopropane (n-propyl bromide) (CAS 106-94-5)? Sicherheit, Regulierung & Perspektive EU/USA/Kanada | ProductGuru
- How to Factory Reset a Google Pixel 8: Step by Step
- Top Dental Care Deals 2026: Save Smart and Smile Brighter
- Top Rated Snacks Brands: A Comprehensive Guide for Health-Conscious Consumers