Recall details
| Product | Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers. |
|---|---|
| Brand / manufacturer | Sysmex America, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1717-2020 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 21 January 2020 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.