ProductGuru
AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →

[Recalled] Stryker Instruments 0400-770-000 T4 Toga, Pullover, Ex-Large Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against conta

recall fda device united-states

Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.

Recall details

ProductStryker Instruments 0400-770-000 T4 Toga, Pullover, Ex-Large Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of mic
Brand / manufacturerStryker Instruments Div. of Stryker Corporation
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-1494-2013
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date23 May 2013

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Stryker Instruments Div. of Stryker Corporation

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.