[Recalled] Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal T
Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.
Recall details
| Product | Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is the |
|---|---|
| Brand / manufacturer | Stryker Howmedica Osteonics Corp. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1210-2013 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 28 March 2013 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Stryker Howmedica Osteonics Corp.