| Product | Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount |
|---|---|
| Brand / manufacturer | IHB OPERATIONS B.V. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1663-2025 |
| Country of notice | Netherlands |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 27 March 2025 |
| EAN / barcode | 10841736116071 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.