[Recalled] Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.

recall fda device united-states

This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.

Recall details

ProductSiemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.
Brand / manufacturerSiemens Healthcare Diagnostics, Inc.
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-2665-2014
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date19 August 2014
EAN / barcode10810210

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to stop executing commands, including uploading validated results to the LIS. The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentraLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Siemens Healthcare Diagnostics, Inc.

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.