[Recalled] Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and |
|---|---|
| Brand / manufacturer | Med Tec Inc |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0671-2018 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 08 November 2017 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Med Tec Inc
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.