[Recalled] Plum A+ Infusion Pump; List Number: 11973; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical car
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Plum A+ Infusion Pump; List Number: 11973; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care |
|---|---|
| Brand / manufacturer | Hospira Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0205-2014 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 05 February 2012 |
| EAN / barcode | 12570148 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →