[Recalled] Phoenix Recon Targeting Arm Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm Product Usage: The Phoenix Femoral Nail System is to be implanted into the femur for alignment, st

recall fda device united-states

Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.

Recall details

ProductPhoenix Recon Targeting Arm Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm Product Usage: The Phoenix Femoral Nail System is to be implanted into the femur for alignment, stabilization and fixation of fractures caused by trauma or disease, and
Brand / manufacturerZimmer Biomet, Inc.
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-0084-2017
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date09 August 2016

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: Old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old revision tissue sleeve has too much clearance with new revision targeting arm. Risks of encountering the affected product include: the correct instrument may not be readily available for the procedure causing the surgeon to complete the procedure by hand; or a delay in surgery greater than 30 minutes may occur while another Recon Targeting Arm or Recon Soft tissue Sleeve is located or while the surgeon completes the procedure by hand.

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Zimmer Biomet, Inc.

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.

Shop Recalled Phoenix Recon Targeting
Search on Amazon ›Find on eBay ›
As an Amazon Associate and eBay Partner, we may earn from qualifying purchases.

Explore by topic

Looking up a product? Use ProductGuru — free barcode & product database with 35M+ products. Search by EAN / UPC barcode.
Was this page helpful?