[Recalled] Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications — Creepage Distance between the two primary phases in the GEO Transformer is 4.5… | FDA 2012-08-
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications |
|---|---|
| Brand / manufacturer | Philips Healthcare Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2384-2012 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 14 August 2012 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Philips Healthcare Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.
- Skincare and Beauty Brands to Avoid in 2025
- Budget-Friendly Petit-Déjeuners: Affordable and Nutritious Morning Meals
- What is 2,2'-dichloro-4,4'-methylenedianiline (CAS 101-14-4)? Safety, Regulation & EU/US/Canada Perspective | ProductGuru
- Was ist 1,2-Benzenedicarboxylic acid, mixed decyl and hexyl and octyl diesters (CAS 68648-93-1)? Sicherheit, Regulierung & Perspektive EU/USA/Kanada | ProductGuru
- Was ist 2-ethylhexyl 10-ethyl-4,4-dioctyl-7-oxo-8-oxa-3,5-dithia-4-stannatetradecanoate (DOTE) (CAS 15571-58-1)? Sicherheit, Regulierung & Perspektive EU/USA/Kanada | ProductGuru
- Ketchup, Mustard, BBQ & Cheese Sauce Expiry and Shelf Life: What You Need to Know