[Recalled] Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system as follows: Azurion 3 M12, Azurion 3 M15, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20 Models: 722063, 722064, 722067
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system as follows: Azurion 3 M12, Azurion 3 M15, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20 Models: 722063, 722064, 722067, 722068, 722078, 722079. Product Usage: The Azurion series (with |
|---|---|
| Brand / manufacturer | Philips North America, LLC |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2205-2019 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 16 July 2019 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.