[Recalled] Panda Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Panda Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a contro |
|---|---|
| Brand / manufacturer | GE Healthcare, LLC |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1847-2019 |
| Country of notice | United States |
| Risk level | Class I |
| Hazard type | recall |
| Notice date | 15 March 2019 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
This Class I recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Stop using the product immediately and follow the authority's remedy (return, repair, or disposal).
Read the full official notice →
Related recalls from GE Healthcare, LLC
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.