[Recalled] "****ORTHO SENSOR**12-123-A Knee Trial for Trithion***1560 Saw grass Corporate Pkwy***4Th floor***Sunrise, FL 33323***www.orthosensor.com***" Rx Only*** Sterile/EO***Made exclusively for Stryker Orthopaedics***. Is the graph
Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.
Recall details
| Product | "****ORTHO SENSOR**12-123-A Knee Trial for Trithion***1560 Saw grass Corporate Pkwy***4Th floor***Sunrise, FL 33323***www.orthosensor.com***" Rx Only*** Sterile/EO***Made exclusively for Stryker Orthopaedics***. Is the graphic user interface software associated with the graphic user interface, |
|---|---|
| Brand / manufacturer | Orthosensor |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0122-2013 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 04 February 2012 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Ortho sensor in Sunrise, FL is recalling the Graphics User Interface Software 1.8 (which was updated to version 1.82, validated on 02/13/12). Recall was initiated due to customer feedback which included the line of site between the knee ba lancer and the Link station was obstructed by OR metal stands or trays or the knee balancer device was placed atop of or near the metal stands or trays, interru
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.