ProductGuru
AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →

[Recalled] OEC 8800 Flexiview designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 8800 is designed to provide fluoroscopic and spot-film imaging of the

recall fda device united-states

The product on this page is subject to an official recall. We summarise the regulator's findings and the recommended action.

Recall details

ProductOEC 8800 Flexiview designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 8800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clin
Brand / manufacturerGE OEC Medical Systems, Inc
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-2027-2012
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date19 November 2007

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from GE OEC Medical Systems, Inc

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.