[Recalled] Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod 10888628001206, NATRELLE Style 10-210 Gel Round
A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.
Recall details
| Product | Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod 10888628001206, NATRELLE Style 10-210 Gel Round Mod 10888628001213, NATRELLE Style 10-240 Gel Round Mod 10888628001220 |
|---|---|
| Brand / manufacturer | Allergan PLC |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0700-2020 |
| Country of notice | United States |
| Risk level | Class III |
| Hazard type | recall |
| Notice date | 06 November 2019 |
| EAN / barcode | 10888628001206 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: The firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-length element string in the UDI data chain of the 2D barcode, which is not formatted to the GS1 standard.
This Class III recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →