[Recalled] MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX TIB INS 6MM — A complaint was received for a discolored All Poly Tibia | FDA 2014-02
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | MOST Options" System (MOST) HINGED TIB INS SZ 1 16MM HINGED TIB INS SZ 4 6MM M HINGED TIB INS SZ 4 9MM M HINGED TIB INS SZ 4 16MM MOST PROX TIB INS 6MM |
|---|---|
| Brand / manufacturer | Zimmer, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1250-2014 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 25 February 2014 |
| EAN / barcode | 502001016 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Zimmer, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.