[Recalled] MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/BP Combo 47, Catalog B1017-417 Microbiology: For use with MicroScan
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/BP Combo 47, Catalog B1017-417 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoi |
|---|---|
| Brand / manufacturer | Beckman Coulter, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1529-2015 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 08 August 2014 |
| EAN / barcode | 10483099 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Technical Support Bulletin issued in February 2013 did not mention performance limitation for Proteus mirabilis and Imp when using the CLSIM100-S20 revised interpretive criteria.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →