[Recalled] Leica Surgical Operating Microscopes, ARveo 8, Part Number 10449063 — During internal testing, Leica Microsystems has become aware of a component cha… | FDA 2022-08-26
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | Leica Surgical Operating Microscopes, ARveo 8, Part Number 10449063 |
|---|---|
| Brand / manufacturer | Leica Microsystems, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1848-2022 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 26 August 2022 |
| EAN / barcode | 30622001 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Leica Microsystems, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.