[Recalled] ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma. — Electrolyte reference reagent, used for cl
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma. |
|---|---|
| Brand / manufacturer | Beckman Coulter Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2136-2020 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 29 January 2020 |
| EAN / barcode | 15099590202491 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →