[Recalled] IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use the IMMULITE¿2000 systems Analyzers- for the
A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.
Recall details
| Product | IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use the IMMULITE¿2000 systems Analyzers- for the quantitative measurement of estradiol in serum, as an aid in the diagno |
|---|---|
| Brand / manufacturer | Siemens Healthcare Diagnostics, Inc |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1243-2016 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 13 January 2016 |
| EAN / barcode | 10381178 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →