Recall details

ProductHIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS12609, RMS62612, RMS62606, RMS12406, RMS22412, RMS12412, RMS42606, RMS62609, RMS52612, RMS42612, RMS32609, RMS22609, RMS2
Brand / manufacturerRepro-Med Systems, Inc.
Categorydevice
Issuing authorityFDA_DEVICE
ReferenceZ-0318-2018
Country of noticeUnited States
Risk levelClass II
Hazard typerecall
Notice date10 March 2016

Why it was recalled

According to the FDA_DEVICE notice, the identified hazard is: On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.

This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.

What to do if you own this product

Check whether your unit is affected and follow the authority's guidance before further use.

Read the full official notice →

Related recalls from Repro-Med Systems, Inc.

This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.

Shop Recalled HIgH Flo Subcutaneous
Search on Amazon ›Find on eBay ›
As an Amazon Associate and eBay Partner, we may earn from qualifying purchases.

Explore by topic

Looking up a product? Use ProductGuru — free barcode & product database with 35M+ products. Search by EAN / UPC barcode.
Was this page helpful?