[Recalled] GE Healthcare Carescape Patient Data Module — GE Healthcare has recently become aware of a potential safety issue to the ECG… | FDA 2014-09-19
Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.
Recall details
| Product | GE Healthcare Carescape Patient Data Module |
|---|---|
| Brand / manufacturer | GE Healthcare, LLC |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0060-2015 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 19 September 2014 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: GE Healthcare has recently become aware of a potential safety issue to the ECG calculations following a disconnect/reconnect cycle with the Patient Data Module, when used with the Carescape Bx50 monitors. If the PDM is quickly disconnected and reconnected (a less than 10 second cycle) from a Carescape Bx50 monitor following a Patient Discharge or during active monitoring, the ECG waveform an
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →