[Recalled] Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S — Due to calibration curves producing a higher activity than routinely expected a… |
A safety authority has issued a recall or warning for this item. The confirmed facts from the source notice are laid out below.
Recall details
| Product | Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S |
|---|---|
| Brand / manufacturer | The Binding Site Group, Ltd. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0274-2022 |
| Country of notice | United Kingdom |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 15 October 2021 |
| EAN / barcode | 05051700006477 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
This Class II recall concerns device products distributed in United Kingdom and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →