[Recalled] Flexible Intubation Fiberscope, Part: 11302BD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Flexible Intubation Fiberscope, Part: 11302BD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube. |
|---|---|
| Brand / manufacturer | Karl Storz Endoscopy |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-2893-2020 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 27 July 2020 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Karl Storz Endoscopy
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.