[Recalled] DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF — Firm received multiple complaints of anesthesia circuit leaks due to cracks in… | FDA 2025-10-10
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF |
|---|---|
| Brand / manufacturer | Medline Industries, LP |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-0446-2026 |
| Country of notice | United States |
| Risk level | Class I |
| Hazard type | recall |
| Notice date | 10 October 2025 |
| EAN / barcode | 40198459055424 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
This Class I recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Stop using the product immediately and follow the authority's remedy (return, repair, or disposal).
Read the full official notice →
Related recalls from Medline Industries, LP
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.