[Recalled] Dynavox T15 Speech Generating Device ( including Batteries) — The battery is integral to the device | FDA 2017-12-14
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Dynavox T15 Speech Generating Device ( including Batteries) |
|---|---|
| Brand / manufacturer | Dynavox Systems Llc |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1305-2018 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 14 December 2017 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Dynavox Systems Llc
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.