[Recalled] Dimension Vista 500 Intelligent Lab System running on software versions V.3.6.1 SP1 or V.3.6.2, Device listing # D011374, All serial numbers and lots are affected. The Dimension Vista System is an in vitro diagnostic device int
Regulators flagged this product in an official safety action. Here is what the notice says and what it means for owners.
Recall details
| Product | Dimension Vista 500 Intelligent Lab System running on software versions V.3.6.1 SP1 or V.3.6.2, Device listing # D011374, All serial numbers and lots are affected. The Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixi |
|---|---|
| Brand / manufacturer | Siemens Healthcare Diagnostics Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1885-2016 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 24 March 2016 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an aliquot probe rinse between sample aspirations when processing tubes in Sample Racks that are front loaded on the Dimension Vista System.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →