[Recalled] Critical Care Ventilator, Catalog Number(s): 8422400: Babylog VN800 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons. — Three separate and unrelate
This is an official product-safety recall record. Below are the verified details from the issuing authority, in plain language.
Recall details
| Product | Critical Care Ventilator, Catalog Number(s): 8422400: Babylog VN800 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons. |
|---|---|
| Brand / manufacturer | Draeger Medical, Inc. |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1693-2021 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 16 April 2021 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Draeger Medical, Inc.
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.