[Recalled] Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT — Potential for the device safety interlock to fail | FDA 2021-03-04
Authorities recorded a safety problem with this product. The details below come directly from the government recall notice.
Recall details
| Product | Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT |
|---|---|
| Brand / manufacturer | Covidien, LP |
| Category | device |
| Issuing authority | FDA_DEVICE |
| Reference | Z-1383-2021 |
| Country of notice | United States |
| Risk level | Class II |
| Hazard type | recall |
| Notice date | 04 March 2021 |
| EAN / barcode | 10884521741843 |
Why it was recalled
According to the FDA_DEVICE notice, the identified hazard is: Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.
This Class II recall concerns device products distributed in United States and was recorded by FDA_DEVICE. Recalls in this category are published so that owners, retailers, and importers can identify affected units and act quickly.
What to do if you own this product
Check whether your unit is affected and follow the authority's guidance before further use.
Read the full official notice →
Related recalls from Covidien, LP
This page indexes a public government recall notice for reference. We do not sell or verify this product; always confirm the current status on the authority's official site.